NSF BioPACIFIC MIP Wiki

A National User Facility Dedicated to Biomaterials

User Tools

Site Tools


led_plate

Differences

This shows you the differences between two versions of the page.

Link to this comparison view

Both sides previous revisionPrevious revision
Next revision
Previous revision
led_plate [2024/10/10 15:49] mbatesled_plate [2025/01/16 19:39] (current) – [Detailed Specifications] jmuruena
Line 19: Line 19:
  
 * **LED technology:** High-intensity LEDs \\ * **LED technology:** High-intensity LEDs \\
-* **Compatible formats:** 96-well plates and 96-well shell vial plates (1 mL scale)\\ +* **Compatible formats:** 96-well plates and 96-well shell vial plates\\ 
-* **Wavelengths available:** 365, 420, 445, and 530 nm\\+* **Wavelengths available:** 365, 420, 445, 470, and 530 nm\\
 * **Power control:** Adjustable power settings \\ * **Power control:** Adjustable power settings \\
 * **Uniformity:** High across all wells\\ * **Uniformity:** High across all wells\\
  
-    + 
 +{{::lumidox_420nm_power.png?400|}} 
 + 
 +**Precise control of LEDs** (Note: There is no individual LED control)\\ 
 +{{::led_control.png?400|}}\\ 
 +GUI for controlling LED array\\ 
 +* Make sure USB from LED controller is connected to COM3 (USB port on left side).  
 +This GUI can be found on BioPACIFIC MIP laptop for 2411. The GUI is called LED power and it is located on the Desktop.\\ 
 + 
 +**time (sec):** duration in seconds that LEDs are on\\ 
 +**current (mA):** intensity of light (range 100 -3,000 mA)\\ 
 +**delay (sec):** time between pulses \\ 
 +**total time (min):** total time of experiment including on and off time\\ 
 + 
 ===== Safety Concerns ===== ===== Safety Concerns =====
  
 The LumiDox Gen II LED Plate produces high-intensity light, which requires the following safety precautions:\\ • Always inspect the LEDs before use to ensure no damage or malfunction.\\ • Do not stare directly at the light sources during operation.\\ • Use protective eyewear rated for specific wavelengths, especially when working with high-intensity light in the UV range.\\ • Ensure that all users are familiar with proper operation procedures and safety protocols outlined by the manufacturer. The LumiDox Gen II LED Plate produces high-intensity light, which requires the following safety precautions:\\ • Always inspect the LEDs before use to ensure no damage or malfunction.\\ • Do not stare directly at the light sources during operation.\\ • Use protective eyewear rated for specific wavelengths, especially when working with high-intensity light in the UV range.\\ • Ensure that all users are familiar with proper operation procedures and safety protocols outlined by the manufacturer.
led_plate.1728575343.txt.gz · Last modified: 2024/10/10 15:49 by mbates