| - | | **Grids to Get:** | + | |**Thickness:** 25 μm \\ **Diameter: |
| * Ultrathin Carbon Film on a Lacey Carbon Support Film | * Ultrathin Carbon Film on a Lacey Carbon Support Film | ||
| Line 25: | Line 27: | ||
| ===== Instructions for Nanoparticle Dropcast TEM sample preparation ===== | ===== Instructions for Nanoparticle Dropcast TEM sample preparation ===== | ||
| - | Things you will need: | + | **Things you will need:** |
| * Ethanol (or water, IPA, or dimethyl carbonate)\\ | * Ethanol (or water, IPA, or dimethyl carbonate)\\ | ||
| Line 35: | Line 37: | ||
| * UltraThin 3 nm Carbon Grids\\ | * UltraThin 3 nm Carbon Grids\\ | ||
| * Glass slide\\ | * Glass slide\\ | ||
| - | 1. Check the MSDS for mixture with ethanol (make sure the nanoparticles are NOT pyrophoric)\\ | + | **1.** Check the MSDS for mixture with ethanol (make sure the nanoparticles are NOT pyrophoric)\\ |
| - | 2. Use the small metal scooper to extract about ½-1 cms worth of nanoparticles\\ | + | **2.** Use the small metal scooper to extract about ½-1 cms worth of nanoparticles\\ |
| - | 3. Place the nanoparticles in a small vial\\ | + | **3.** Place the nanoparticles in a small vial\\ |
| - | 4. Fill the small vial with ethanol up to the 1 cm below the rim\\ | + | **4.** Fill the small vial with ethanol up to the 1 cm below the rim\\ |
| - | 5. Sonicate the filled small vial for 10 minutes on high power (make sure to find an anti-node)\\ | + | **5.** Sonicate the filled small vial for 10 minutes on high power (make sure to find an anti-node)\\ |
| - | 6. Centrifuge for 20 minute (make sure to balance the centrifuge)\\ | + | **6.** Centrifuge for 20 minute (make sure to balance the centrifuge)\\ |
| - | 7. Use a pipette to collect several uL of fluid just above the concentrated part at the bottom of the vial. See the arrow below . The pipette should just touch above dark part.\\ | + | **7.** Use a pipette to collect several uL of fluid just above the concentrated part at the bottom of the vial. See the arrow below . The pipette should just touch above dark part\\ |
| {{ : | {{ : | ||
| - | 8. Deposit 1 uL of fluid onto the dull (not shiny) side of the grid.\\ | + | **8.** Deposit 1 uL of fluid onto the dull (not shiny) side of the grid\\ |
| - | 9. Let the grid dry on a glass slide.\\ | + | **9.** Let the grid dry on a glass slide\\ |
| - | 10. Vacuum bake at 50-400 degree C for several hours if contamination is found of the surface.\\ | + | **10.** Vacuum bake at 50-400 degree C for several hours if contamination is found of the surface\\ |
| - | 11. Plasma clean as needed.\\ | + | **11.** Plasma clean as needed\\ |