- | ===== Labconco Lyophilizer | + | ===== Labconco Lyophilizer |
+ | The Labconco lyophilizer is used for freeze-drying chemical samples. Before lyophilization, | ||
- | The Labconco | + | **1. Wake the System**\\ |
- | + | The lyophilizer | |
- | **Procedure: | + | |
- | -**Wake the System:** | + | **2. Turn on the Collector**\\ |
- | The lyophilizer will start in Sleep mode. Tap the screen to wake it up. | + | On the home screen, select the Collector tab to turn on the collector. The system will begin pumping down to -84°C, which takes approximately 10 minutes. |
- | -**Turn on the Collector:** | + | **3. Drain the Collector |
- | On the home screen, select | + | If prompted to " |
+ | - Insert | ||
+ | - Attach the waste tube to the collector outlet on the left side of the instrument. | ||
+ | | ||
- | Drain the Collector (if needed): | + | **4. Check the Tower and Vacuum Ports**\\ |
- | If prompted to "Drain the Collector" | + | Ensure |
- | Insert the waste tube into a solvent waste bottle. | + | |
- | Attach | + | |
- | Once the solvent has drained, detach the waste line by pressing the detachment button at the port. (Failure to detach the waste line will prevent | + | |
- | Check the Tower and Vacuum | + | **5. Start the Vacuum**\\ |
- | | + | Go to the Vacuum tab and set the Vacuum Set Point to 0.01 mbar. Click Start. Turn on the secondary pump located below the instrument by pressing the green button. The pump should engage, and the system will begin to pump down. If the vacuum |
- | Start the Vacuum: | + | **6. Attach Your Sample**\\ |
- | Go to the Vacuum tab and set the Vacuum Set Point to 0.01 mbar. | + | Once your sample is properly frozen, place it into a vacuum jar and attach it to a vacuum port. Slowly open the vacuum port by turning |
- | Click Start. | + | |
- | Turn on the secondary pump located below the instrument | + | |
- | If the vacuum reading stays " | + | |
- | Attach Your Sample: | + | **7. Remove the Sample**\\ |
- | Once your sample | + | After freeze-drying |
- | Slowly open the vacuum port by turning the black knob. Be cautious: if the vacuum is opened too quickly, or if the sample | + | |
- | Remove | + | **8. Shut Down the System**\\ |
- | After freeze-drying is complete, close the vacuum port and remove your sample. | + | Turn off the condenser |
+ | | ||
- | Shut Down the System: | ||
- | Turn off the condenser and vacuum by selecting " | ||
---- | ---- | ||