saxs-intro
Differences
This shows you the differences between two versions of the page.
Next revision | Previous revision | ||
saxs-intro [2024/10/14 00:51] – created phillipkohl | saxs-intro [2024/10/14 04:13] (current) – phillipkohl | ||
---|---|---|---|
Line 1: | Line 1: | ||
+ | {{: | ||
+ | |||
+ | **About the SAXS/WAXS Diffractometer** | ||
+ | |||
Small angle X-ray scattering (SAXS) is a premier technique for determining structural characteristics on the nanometer scale. The utilization of SAXS for high throughput sample characterization in a laboratory environment however is typically inhibited by a variety of factors, such as low x-ray brilliance for in-house setups, lack of synchrotron availability, | Small angle X-ray scattering (SAXS) is a premier technique for determining structural characteristics on the nanometer scale. The utilization of SAXS for high throughput sample characterization in a laboratory environment however is typically inhibited by a variety of factors, such as low x-ray brilliance for in-house setups, lack of synchrotron availability, | ||
Line 6: | Line 10: | ||
To optimize the workflow, a custom-developed graphical user interface (GUI) has been developed to provide rapid measurement turn around, boasting automated sample alignment, real-time viewing of 2D SAXS data alongside on the fly 1D data reduction, as well as AI driven automated data analysis. | To optimize the workflow, a custom-developed graphical user interface (GUI) has been developed to provide rapid measurement turn around, boasting automated sample alignment, real-time viewing of 2D SAXS data alongside on the fly 1D data reduction, as well as AI driven automated data analysis. | ||
+ | |||
+ | **Diffractometer Information** | ||
+ | |||
+ | * Incident wavelength: 1.33A | ||
+ | * Default beam size: 800 x 800 um for transmission, | ||
+ | * Minimum beam size: 200 x 200 um | ||
+ | * Possible q range: 0.002-2.5 A^-1 | ||
+ | * Possible sample to detector distances: 140-2400 mm | ||
+ | * Possible temperature range for oven: 0-200 C | ||
+ | * Detector pixel size: 75 x 75 um | ||
+ | * Source: Excillum Metal-jet D2+ | ||
+ | * Detector: Eiger2 4M | ||
+ | |||
+ | **Shipping Information** | ||
+ | |||
+ | Mail-in samples to be run should be sent to either Phillip Kohl or Youli Li using the following address: | ||
+ | |||
+ | Materials Research Laboratory, MC 5121 | ||
+ | University of California | ||
+ | Santa Barbara, CA 93106-5121 | ||
+ | (805) 893-7233 | ||
+ | |||
+ | **How to run a sample** | ||
+ | |||
+ | **Step 1:** Load samples into the desired sample holder and slide the sample holder onto the base of magnetic base of the sample stage. | ||
+ | |||
+ | {{: | ||
+ | |||
+ | **Step 2:** Close the hutch and secure it by fastening the top and bottom clamps of the SAXS hutch. | ||
+ | |||
+ | {{: | ||
+ | |||
+ | **Step 3:** On the SAXS computer check to see that the GUI is open. If not, or if the computer has to be restarted open the file named "TO REOPEN GUI" located on the computer' | ||
+ | |||
+ | **Step 4:** Check to see that the system at Full Power if at Low Power Mode, click on the switch to power up the source. If you are the first user of the day, or if you changed the beam size, click on the 'align optics' | ||
+ | |||
+ | {{: | ||
+ | |||
+ | **Step 5:** Go to the ' | ||
+ | |||
+ | {{: | ||
+ | |||
+ | **Step 6:** If your sample is smaller than 2 mm, you will likely need to use the x-ray beam to accurately align your sample. To do so, fill in the necessary boxes in the 'Scan Parameters' | ||
+ | |||
+ | {{: | ||
+ | |||
+ | **Step 7:** When the scan is complete look at the updated graph in the middle of the GUI and click on the point where your sample is. For capillary samples, you will want to click on the center of the " | ||
+ | |||
+ | **Step 8:** Fill in the desired scan range (qmin and qmax), exposure time, and file name for the experiment. Note: To avoid errors, keep the length of the file name below 28 characters (you can change the file name after it is done running). | ||
+ | |||
+ | {{: | ||
+ | |||
+ | **Step 8b:** If you are running a temperature experiment, drop down the temperature experiment panel, and click on the switch. Fill in the targeted temperature for the experiment, and how long to for the computer to wait before running the exposure upon reaching the targeted temperature. | ||
+ | |||
+ | {{: | ||
+ | |||
+ | **Step 8c:** If you are running a sample in the GI geometry, drop down the grazing incidence experiment panel, and click on the switch. In the ' | ||
+ | |||
+ | {{: | ||
+ | |||
+ | **Step 8:** When all relevant information for the experiment have been entered click on the ' | ||
+ | |||
+ | **Step 9:** Repeat Steps 5-8 for all samples that you would like to run. Once complete click on the 'Run Samples' | ||
+ | |||
+ | |||
saxs-intro.1728867098.txt.gz · Last modified: 2024/10/14 00:51 by phillipkohl