====== Sample Preparation ====== {{:workflow_diagram_planning_whiteboard_in_green_yellow_modern_professional_style_1_.png?600 |}} ===== Grids ===== |**Thickness:** 25 μm \\ **Diameter:** 3 mm \\ Film on “dull” side\\ “Shiny” side has the metal only \\ | * Ultrathin Carbon Film on a Lacey Carbon Support Film * https://www.tedpella.com/Support_Films_html/Support_Films_and_Substrates_Overview.aspx * 01824 Ultrathin Carbon Film on Lacey Carbon Support Film, 400 mesh, Copper * 160 PELCO®️ TEM Grid Storage Box {{ :tem_grids.jpg?600 |}} ---- ===== Basic Shake and Bake ===== Put a small amount of powder in Eppendorf tub (enough to coat the bottom). Put one grid into the tube. Close the tube and shake for 1 minute. Open the tube and remove the grid. The grid should have enough material on it and is ready to be imaged in the TEM. ---- ===== Instructions for Nanoparticle Dropcast TEM sample preparation ===== **Things you will need:** * Ethanol (or water, IPA, or dimethyl carbonate)\\ * A small metal scooper\\ * A sonicator\\ * Several small vials that are centrifuge compatible\\ * A centrifuge\\ * Nanoparticles\\ * UltraThin 3 nm Carbon Grids\\ * Glass slide\\ **1.** Check the MSDS for mixture with ethanol (make sure the nanoparticles are NOT pyrophoric)\\ **2.** Use the small metal scooper to extract about ½-1 cms worth of nanoparticles\\ **3.** Place the nanoparticles in a small vial\\ **4.** Fill the small vial with ethanol up to the 1 cm below the rim\\ **5.** Sonicate the filled small vial for 10 minutes on high power (make sure to find an anti-node)\\ **6.** Centrifuge for 20 minute (make sure to balance the centrifuge)\\ **7.** Use a pipette to collect several uL of fluid just above the concentrated part at the bottom of the vial. See the arrow below . The pipette should just touch above dark part\\ {{ :vial2.png?200 |}} **8.** Deposit 1 uL of fluid onto the dull (not shiny) side of the grid\\ **9.** Let the grid dry on a glass slide\\ **10.** Vacuum bake at 50-400 degree C for several hours if contamination is found of the surface\\ **11.** Plasma clean as needed\\